  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 32
Table
2. DIP
Switch Positions
to Use for Different
Ranges
SW1
Range /
Position
99.9
ohms
999
ohms
9.99K
ohms
99.9K
ohms
999K
ohms
1
ON OFF OFF OFF OFF
2 OFF ON OFF OFF OFF
3 OFF OFF ON OFF OFF
4
OFF OFF OFF ON OFF
5 OFF OFF OFF OFF ON
6 ON ON OFF OFF OFF
7
OFF OFF
ON ON ON
8 OFF OFF OFF OFF OFF
SW2
Range /
Position
99.9
ohms
999
ohms
9.99K
ohms
99.9K
ohms
999K
ohms
1
ON
I
OFF
OFF OFF
OFF
2
OFF ON
OFF OFF
OFF
3 OFF
OFF ON
OFF OFF
4
OFF
OFF OFF ON
OFF
5 OFF
OFF OFF OFF
ON
6
OFF OFF
ON OFF
OFF
7
ON
OFF OFF ON
OFF
8 OFF
ON OFF
OFF
ON
the Digi -Ohm module
onto the
rear
of the DMS as shown
in Fig. 7.
Connect
test leads to TBI at TP
+
and TP - . Connect an appropriate
power source
to TP+ and GND and
turn on the power supply.
At this
point,
your DMS module should be
displaying "EEE," indicating an
overrange (infinite- resistance) con-
dition. Shorting
TP +
to
TP -
should
cause
the
displayed
reading to drop
to
"000,"
indicating a 0 -ohm (short -
circuit) condition.
If you
encounter
any problems at
this stage,
UI, CR/ or the headers
are the
probable cause.
There are two
ways
to calibrate the
Digi -Ohm module. The first is
to use
several precision resistors
of differ-
ent
values,
one per range. The
sec-
ond, more accurate, method is to use
a precision current meter.
If you
use precision
resistors
for
calibration, select
resistors
of
known
value
in
the
middle
of each
range
to
be calibrated. If possible, the resis-
tors should be 0.1- percent tolerance.
For the second method, the
current
Fig.
6.
Actual
-size etching -and-
drilling guide
(A) and
wiring
diaram
(B) for LED
boards.
meter used should
have a rated ac-
curacy
of 0.1
percent.
If
you
are
using the
DIP -switch
method of
range selection,
refer to
Table 2 and position
S/ and S2
for
the lowest
range
(99.9
ohms
full -
scale).
If you
are using the
remote
switch
panel, connect
the cables at
S/
and S2
and set the selector
switch to
the 100 -ohm
range.
When
using
the
fixed -range arrangement,
substitute
the appropriate
values
(see
Tables
1
and 2)
for
the
specific
fixed range
you have selected.
For calibration
with
precision
re-
sistors,
connect
a 50 -ohm test resis-
tor
between
TP +
and
TP -
and adjust
the setting of
R2
until the
DMS
dis-
plays "050." For calibration
with
a
current meter, set the
milliammeter
to a
range
that
will measure 10 mA.
Adjust the
setting
of R2 for a
reading
of
10.00 mA on the milliammeter.
Repeat
the
procedure
for
each
range,
using the proper
middle
-of-
the -range
resistor (first method)
or adjusting
for the current
indicated in Table 1.
The appropriate decimal
point and
panel
LED (12
or
KSI)
should
light as
each
range is
selected.
If you
encounter
any difficulties,
check to see if the proper
value
of
trimmer potentiometer has been in-
stalled for each range. If
no
current
is
present, check
U1, CR1 and the
range -selection arrangement.
Make
any
necessary repairs and then
repeat
the calibration
procedure.
After
calibrating
all
ranges,
apply
a small amount of fingernail enamel
to the trimmer screws to prevent
Say You Saw
It
In Modern
Electronics February
1988
/
MODERN ELECTRONICS
/ 43.